Supermini Application Byte

 

Supermini benchtop WDXRF meets performance requirements of ASTM C114 standard test method for hydraulic cement

In a strictly regulated industry, cement producers must maintain manufacturing processes that are accurate, running efficiently, and thus ensure that their customers receive the highest quality material ideally suited for a specific application. So a cement plant must continuously analyze its products to meet these requirements. The ability to accurately analyze for the major components as well as for minute concentrations of light elements (as low as sodium) has recently been demonstrated on the Rigaku Supermini 200 W benchtop WDXRF spectrometer. This small unit uses a 200 W Pd X-ray tube, has a 12-position sample carousel for automated analysis, and an air-cooled X-ray tube, which eliminates the need for cooling water or a chiller.

The performance goals met by the Supermini are those set forth in ASTM C114 : Standard Test Methods for Chemical Analysis of Hydraulic Cement. Overall C114 is a measure of both the analyzing system's stability and the user's ability at sample preparation—the sample must be repeatedly produced as a pressed pellet or fused disk and show only very small analytical variations, the assumption being that if both instrument stability and sample preparation are optimized then a linear calibration will be achieved using reference materials such as NIST Portland Cements and both accuracy and precision will be within the ASTM method specifications. Reference materials used consisted of the NIST Certified materials 1880a, 1881a, 1884a, 1885a, 1886a, 1887a, 1888a, 1889a.

The initial test was performed using fused disks supplied by the Arkansas Department of Transportation. The fusion recipe consisted of a 1:10 dilution of sample into a 50:50 lithium salt flux. An analytical routine was created on the Supermini using the NIST Certified Reference Materials (CRMs). One of the CRMs (1884a), prepared in duplicate, was run as an unknown to determine if the Supermini could comply with C114.

The following tables demonstrate the results for these runs in accordance to C114 specifications.

Component

Range (mass%)

Component

Range (mass%)

CaO

57.5867.87

TiO2

0.0840.366

SiO2

18.63722.38

Na2O

0.0211.068

Al2O3

3.8757.06

P2O5

0.0220.306

Fe2O3

0.1523.09

Mn2O3

0.0070.259

SO3

2.0864.622

Cl

0.00190.013

MgO

0.8144.475

ZnO

0.0010.107

K2O

0.0931.228

 

 

Table 1: Calibration range for test materials

Standard Error (SE) results: 

Element

Na2O

MgO

Al2O3

SiO2

P2O5

SO3

%

0.021

0.043

0.035

0.107

0.005

0.041

 

 

 

 

 

 

 

Element

K2O

CaO

TiO2

Mn2O3

Fe2O3

ZnO

%

0.008

0.225

0.008

0.001

0.036

0.001

Table 2: Calibration Standard Error

Note: SE is a measure of the goodness of a specific calibration (as shown in the equation below),  the average difference (in unit concentration) between the suggested standard values and their proximity to calculated calibration line.

where σest is the SE, Y is the value of a calibration standard, Y' is measured value of a calibration standard, and N is the number of calibration standards employed.

Each of the duplicate samples were run 10x on the Supermini. All tests cycled the disk in and out of the XRF system between analysis runs.

Component

Certified

Average and standard deviation — 10x repeat measurement

Duplicate pellet difference

Difference between averaged results of duplicate disks and certified value

Disk 1

Disk 2

Average

Standard

deviation

Average

Standard deviation

ASTM requirement

Result

ASTM requirement

Result

Na2O

0.216

0.202

0.026

0.184

0.013

0.03

0.018

0.05

0.023

MgO

4.475

4.487

0.013

4.453

0.018

0.16

0.034

0.2

0.005

Al2O3

4.26

4.29

0.017

4.24

0.016

0.20

0.050

0.2

0.001

SiO2

20.57

20.59

0.015

20.53

0.028

0.16

0.054

0.2

0.010

P2O5

0.128

0.134

0.002

0.116

0.004

0.03

0.018

0.03

0.003

SO3

2.921

2.907

0.010

2.88

0.009

0.10

0.026

0.1

0.028

K2O

0.997

0.999

0.005

1.000

0.006

0.03

0.001

0.05

0.003

CaO

62.26

62.12

0.029

62.24

0.030

0.20

0.110

0.3

0.079

TiO2

0.186

0.184

0.009

0.181

0.006

0.02

0.003

0.03

0.004

Mn2O3

0.085

0.086

0.003

0.087

0.003

0.03

0.001

0.03

0.002

Fe2O3

2.695

2.653

0.010

2.638

0.009

0.10

0.015

0.1

0.049

ZnO

0.010

0.010

0.000

0.010

0.001

0.03

0.000

0.03

0.000

Table 3: Repeat results (sample CRM 1884a) Unit mass% 

Conclusion:

ASTM C114 requirements can easily be achieved using this benchtop WDXRF system. Benefits from the Supermini obviously are efficient use of space and comparative ease of operation. These benefits, plus minimum service requirements, lower initial financial outlay and lower ongoing cost of ownership, make the Supermini an ideal solution for cement analysis by XRF.

 

Tags: XRF, Supermini, high-power WDXRF